CALCulate Subsystem 3
N1911A/1912A P-Series Power Meters Programming Guide 197
CALCulate[1]|2|3|4:RELative:STATe <boolean>
This command enables/disables relative mode. If the command is:
• disabled, the measurement signal remains unchanged.
• enabled, the current relative value set by
CALCulate:RELative:MAGnitude:AUTO is applied to the measurement
signal.
Syntax
Example
Reset Condition
On reset, relative mode is disabled.
Query
CALCulate[1]|2|3|4:RELative:STATe?
CALC1:REL:STAT OFF This command disables the relative mode
on the upper window/upper
measurement.
:STAT
:REL
0|OFF
1|ON
Space
?
CALC
1
2
3
4