EasyManuals Logo

Agilent Technologies N1911A Programming Guide

Agilent Technologies N1911A
803 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #178 background imageLoading...
Page #178 background image
2 MEASurement Commands
150 N1911A/1912A P-Series Power Meters Programming Guide
MEASure[1]|2|3|4[:SCALar][:POWer:AC]:RELative?
[<expected_value>[,<resolution>[,<source list>]]]
This command sets the specified window’s measurement function to single
channel with relative mode on, aborts, configures then initiates the
specified channel, calculates the measurement result and places the result
on the bus. The result is a ratio based measurement and is expressed in
the units defined by UNIT[1]|2|3|4:POWer:RATio. The relative value
used is that set by the CALCulate:RELative:MAGNitude:AUTO command.
Syntax
Parameters
Refer to “Optional Parameters” on page 90 for additional details on the
parameters in this command.
:POW
:AC
?
:SCAL
:REL
Space
expected_value
DEF
resolution
,
DEF
source list
,
MEAS
1
2
3
4
Item Description/Default Range of Values
expected_value
(for the expected
power level)
A numeric value for the expected power level.
The units of measurement are dBm and W. The
default units are defined by UNIT:POWer.
sensor dependent
DEF
1
resolution A numeric value for the resolution. If
unspecified the current resolution setting is
used.
1 to 4
2
1.0, 0.1, 0.01, 0.001
DEF
1

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals