SERVice Subsystem 15
N1911A/1912A P-Series Power Meters Programming Guide 723
SERVice:SENSor[1]|2:POWer:USABle:MINimum?
This query returns the minimum power that can be accurately measured
by the currently connected sensor. It is applicable to E- Series sensors
only. Maximum power information is stored in the sensor’s EEPROM.
Syntax
Example
Error Messages
If no sensor is connected, error –241, “Hardware missing” occurs.
If a a sensor other than an E- Series sensor is connected, error –241
“Hardware missing” occurs.
If the E- Series sensor currently connected does not contain the
necessary information in EEPROM, error –241 “Hardware missing”
occurs.
SERV:SENS:POW:USAB:MIN? This query returns the minimum power
that can be accurately measured by the
E- Series sensor currently connected to
Channel A.
:POW
?
:USAB
:MIN
SERV
1
2
:SENS