MEMory Subsystem 7
N1911A/1912A P-Series Power Meters Programming Guide 279
MEMory:TABLe:FREQuency
<numeric_value>{,<numeric_value>}
This command is used to enter frequency data into the current selected
table. Any previous frequency list is cleared before the new frequency list
is stored. The frequencies must be entered in ascending order. Entries in
the frequency lists correspond as shown in Table 7- 21 with entries in the
calibration/offset factor lists.
NOTE
For sensor calibration tables only, the first calibration factor entered using the
MEMory:TABLe:GAIN command is used as the reference calibration factor.
NOTE
Predefined sensor calibration table is not applicable for N8480 Series power sensors with
Option CFT. Therefore you are required to create a new sensor calibration table for the
sensors when a sensor calibration table is needed.