10 SENSe Subsystem
430 N1911A/1912A P-Series Power Meters Programming Guide
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO <boolean>
This command is only valid when used with an E-Series power sensor or
N8480 Series power sensor (excluding Option CFT). Its purpose is to
enable and disable autoranging. When autoranging is ON, the power meter
selects the best measuring range for the measurement. When autoranging
is set to OFF, the power meter remains in the currently set range.
The [SENSe[1]]|SENSe2:POWer:AC:RANGe command disables autoranging.
If INITiate:CONTinuous is set to ON and TRIGger:SOURce is set to
IMMediate, the range tracks the input power if
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO is ON.
If the power meter is not making measurements then autoranging only
occurs when the power meter is triggered.
Syntax
Example
NOTE
For E9320 power sensor, only UPPER and LOWER ranges are available in Normal and
Triggered modes.
POW:AC:RANG:AUTO 0 This command disables autoranging.
SENS
1 :AC
POW
SENS2
:
:AUTO
0|OFF
1|ON
Space
?
:RANG

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals