7MEMory Subsystem
268 N1911A/1912A P-Series Power Meters Programming Guide
MEMory:CLEar:TABLe
This command is used to clear the contents of the table currently selected
using MEMory:TABLe:SELect. Although the table remains, a
MEMory:TABLe:FREQuency|GAIN:POINts? query returns a 0 as the table
contents are empty.
This command is an alternative form of the MEMory:CLEar[:NAME]
command. The difference is the method in which the table is selected.
Syntax
Example
Error Message
If no table is selected, error –221, “Settings conflict” occurs.
NOTE
The contents cleared using this command are non-recoverable.
MEM:CLE:TABL This command clears the contents of the
currently selected table.