1 Power Meter Remote Operation
36 N1911A/1912A P-Series Power Meters Programming Guide
Figure 1-3 Averaging Range Hysteresis
Filter Length
You specify the filter length using the following command:
[SENSe[1]]|SENSe2:AVERage:COUNt <numeric_value>
The range of values for the filter length is 1 to 1024. Specifying this
command disables automatic filter length selection. Increasing the value of
the filter length reduces measurement noise. However, the time to take the
measurement is increased.