CALibration Subsystem 4
N1911A/1912A P-Series Power Meters Programming Guide 215
CALibration[1]|2:ZERO:NORMal:AUTO <boolean>
This command provides a quick way of zeroing the NORMAL path of an
E9320 Series sensor. The average only path is unaffected. This command
can only be used to zero an E9320 Series sensor and a P- Series sensor.
The command causes the power meter to perform its zeroing routine, on
the specified channel, when ONCE is selected. This adjusts the power meter
for a zero power reading with no power supplied to the power sensor.
The 0|OFF parameter is only required for the query response and is
ignored in the command. If 1|ON is selected on an E9320 Series sensor, it
causes the error –224, “Illegal parameter value” to occur.
Except when using a P- Series sensor, this command assumes that the
E9320 Series sensor is not connected to a power source.
Syntax
Example
NOTE
The P-Series sensor only has a NORMAL path. Hence, the reason this E9320 Series sensor
command is allowed to function.
CAL2:ZERO:NORM:AUTO ONCE This command causes the power meter to
perform a zeroing routine on Channel B.
:AUTO:NORM
:ZERO
CAL
1
2
0|OFF
Space
?
ONCE

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals