SERVice Subsystem 15
N1911A/1912A P-Series Power Meters Programming Guide 715
SERVice:SENSor[1]|2:CPLace?
This query returns the calibration place in P- Series, E- Series sensors and
N8480 Series sensors. Calibration place information is stored in the
sensor’s EEPROM.
Syntax
Example
Error Messages
• If no power sensor is connected, error –241 “Hardware missing” occurs.
• If a sensor other than a P- Series, N8480 Series or E- Series power
sensor is connected, error –241 “Hardware missing” occurs.
SERV:SENS2:CPL? This query returns the place of
calibration of the P- Series, E- Series
sensor or N8480 Series sensor connected
to Channel B.