SERVice Subsystem 15
N1911A/1912A P-Series Power Meters Programming Guide 715
SERVice:SENSor[1]|2:CPLace?
This query returns the calibration place in P- Series, E- Series sensors and
N8480 Series sensors. Calibration place information is stored in the
sensor’s EEPROM.
Syntax
Example
Error Messages
If no power sensor is connected, error –241 “Hardware missing” occurs.
If a sensor other than a P- Series, N8480 Series or E- Series power
sensor is connected, error –241 “Hardware missing” occurs.
SERV:SENS2:CPL? This query returns the place of
calibration of the P- Series, E- Series
sensor or N8480 Series sensor connected
to Channel B.
:CPL
?
SERV
1
2
:SENS