Power Meter Remote Operation 1
N1911A/1912A P-Series Power Meters Programming Guide 37
Range
The power meter has no internal ranges which can be set. The only ranges
that can be set are those of the E- Series power sensor and N8480 Series
power sensors (excluding Option CFT). With an E- Series power sensor or
N8480 Series power sensors (excluding Option CFT), the range can be set
either automatically or manually. Use autoranging when you are not sure
of the power level you will be measuring.
Setting the Range
To set the range manually use the following command:
[SENSe[1]]|SENSe2:POWer:AC:RANGe <numeric_value>
If the <numeric_value> is set to:
• 0, the sensor’s lower range is selected. (For example, this range is –70
to –13.5 dBm for the E4412A power sensor.)
• 1, the sensor’s upper range is selected. (For example, this range is
–14.5 to +20 dBm for the E4412A power sensor.)
For details on the range limits of other E- Series power sensor and N8480
Series power sensor (excluding Option CFT), refer to the appropriate
power sensor manual.
For further information on this command refer to page 428.
To enable autoranging use the following command:
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO ON
Use autoranging when you are not sure of the power level you will be
measuring.

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals