Power Meter Remote Operation 1
N1911A/1912A P-Series Power Meters Programming Guide 17
Using the CONFigure Command
When you execute this command, the power meter presets the optimum
settings for the requested configuration (like the MEASure? query).
However, the measurement is not automatically started and you can
change measurement parameters before making measurements. This allows
you to change the power meter’s configuration from the preset conditions.
The power meter offers a variety of low- level commands in the SENSe,
CALCulate, and TRIGger subsystems. For example, if you want to change
the averaging use the [SENSe[1]]|SENSe2:AVERage:COUNt command.
Use the INITiate or READ? query to initiate the measurement.
Using READ?
CONFigure does not take the measurement. One method of obtaining a
result is to use the READ? query. The READ? query takes the measurement
using the parameters set by the CONFigure command then sends the
reading to the output buffer. Using the READ? query obtains new data.
Using INITiate and FETCh?
CONFigure does not take the measurement. One method of obtaining the
result is to use the INITiate and FETCh? commands. The INITiate
command causes the measurement to be taken. The FETCh? query
retrieves a reading when the measurement is complete, and sends the
reading to the output buffer. FETCh? can be used to display the
measurement results in a number of different formats (for example, A/B
and B/A) without taking fresh data for each measurement.
CONFigure Examples
The following program segments show how to use the commands READ?,
INITiate and FETCh? and CONFigure to make measurements.
It is advisable to read through these examples in order as they become
increasingly more detailed.
These examples give an overview of the CONFigure command. For further
information on the CONFigure commands refer to Chapter 2,
“MEASurement Commands”.

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals