7MEMory Subsystem
290 N1911A/1912A P-Series Power Meters Programming Guide
MEMory:TABLe:SELect <character_data>
This command is used to activate either a sensor calibration table (8480
Series sensors and N8480 Series sensors with Option CFT only), or a
frequency dependent offset table. A table must be activated before any
operation can be performed on it.
Syntax
Parameters
Example
Query
MEMory:TABLe:SELect?
The query returns the name of the currently selected table.
Item Description/Default Range of Values
character_data Details the new table name. A maximum of
12 characters can be used.
A to Z (uppercase)
a to z (lowercase)
0 - 9
_ (underscore)
MEM:TABL:SEL "Sensor1" This command selects a sensor calibration
table named “Sensor1”.
MEM
:TABL
:SEL
?
Space
character_data