EasyManuals Logo

Agilent Technologies N1911A Programming Guide

Agilent Technologies N1911A
803 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #318 background imageLoading...
Page #318 background image
7MEMory Subsystem
290 N1911A/1912A P-Series Power Meters Programming Guide
MEMory:TABLe:SELect <character_data>
This command is used to activate either a sensor calibration table (8480
Series sensors and N8480 Series sensors with Option CFT only), or a
frequency dependent offset table. A table must be activated before any
operation can be performed on it.
Syntax
Parameters
Example
Query
MEMory:TABLe:SELect?
The query returns the name of the currently selected table.
Item Description/Default Range of Values
character_data Details the new table name. A maximum of
12 characters can be used.
A to Z (uppercase)
a to z (lowercase)
0 - 9
_ (underscore)
MEM:TABL:SEL "Sensor1" This command selects a sensor calibration
table named “Sensor1”.
MEM
:TABL
:SEL
?
Space