9 PSTatistic Subsystem
336 N1911A/1912A P-Series Power Meters Programming Guide
PSTatistic[1]|2:CCDF:DATa:MAX <numeric_value>
This command is used to set the maximum value of X-axis CCDF trace.
Syntax
Parameters
Example
Reset Condition
On reset, the maximum value for CCDF trace X- axis is set to 50 dB.
PST
2
1
:CCDF
:DAT
:MAX
numeric_value
?
Item Description/Default Range of Values
numeric_value X-axis CCDF trace maximum value
in dB.
• Minimum value: 5.00 dB
• Maximum value: 50.00 dB
5.00 to 50.00
PST1:CCDF:DAT:MAX 10 This command sets the maximum value of
X- axis CCDF trace to 10 dB.