EasyManuals Logo

Agilent Technologies N1911A Programming Guide

Agilent Technologies N1911A
803 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #241 background imageLoading...
Page #241 background image
CALibration Subsystem 4
N1911A/1912A P-Series Power Meters Programming Guide 213
CALibration[1]|2:ZERO:AUTO [ONCE|ON|OFF|0|1]
This command causes the power meter to perform its zeroing routine on
the specified channel when enabled. This adjusts the power meter for a
zero power reading with no power supplied to the power sensor.
1|ON can only be used with a P-Series sensor. When 1|ON is enabled the
the zero is maintained by a combination of on- the- fly zero measurements
and temperature compensation.
The 0|OFF parameter is only required for the query response and is
ignored in the command.
Except when using a P- Series sensor, this command assumes that a power
sensor is not connected to a power source.
Syntax
Example
Reset Condition
On reset, automatic zeroing is disabled.
CAL2:ZERO:AUTO ONCE This command causes the power meter to
perform a zeroing routine on Channel B.
:AUTO
:ZERO
0|OFF
1|OFF
Space
?
CAL
1
2
ONCE

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals