CALibration Subsystem 4
N1911A/1912A P-Series Power Meters Programming Guide 211
CALibration[1]|2:RCFactor <numeric_value>
This command is used with 8480 Series power sensors or N8480 Series
power sensors with Option CFT to set the reference calibration factor of
the specified channel. Reference calibration factors can also be set using
sensor calibration tables. The power meter uses the most recently set
reference calibration factor.
Syntax
Parameters
Item Description/Default Range of Values
numeric_value A numeric value:
DEF: the default is 100 %
MIN: 1 %
MAX:150 %
1.0 to 150.0 PCT
DEF
MIN
MAX
:RCF