2 MEASurement Commands
102 N1911A/1912A P-Series Power Meters Programming Guide
CONFigure[1]|2|3|4[:SCALar][:POWer:AC]:DIFFerence
[<expected_value>[,<resolution>[,<source list>]]]
This command sets the measurement function and resolution of the
specified window. It sets the measurement function to difference with
relative mode off.
Syntax
Parameters
Refer to “Optional Parameters” on page 90 for additional details on the
parameters in this command.
:POW
:AC
:SCAL
:DIFF
Space
expected_value
DEF
resolution
,
DEF
source list
,
CONF
1
2
3
4
Item Description/Default Range of Values
expected_value
(for the expected
power level)
The power meter ignores the numeric
value entered in this parameter. Any
value entered is treated like DEF.
sensor dependent
DEF
1
resolution A numeric value for the resolution. If
unspecified the current resolution setting
is used.
1 to 4
2
1.0, 0.1, 0.01, 0.001
DEF
1

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals