11 STATus Subsystem
466 N1911A/1912A P-Series Power Meters Programming Guide
STATus Subsystem
The STATus command subsystem enables you to examine the status of the
power meter by monitoring the following status registers:
• Device status register
• Operation status register
• Questionable status register
The contents of these and other registers in the power meter are
determined by one or more status registers.
Table 11- 25 summarizes the effects of various commands and events on
these status registers:
Tab le 11-25Commands and events affecting Status Register
The contents of the status registers are examined using the following
status register set commands:
:CONDition?
:ENABle <NRf>|<non-decimal numeric>
[:EVENt?]
:NTRansition <NRf>|<non-decimal numeric>
:PTRansition <NRf>|<non-decimal numeric>
Each of these can be used to examine any of the following eleven status
registers:
Status Register *RST *CLS Power On STATus:
PRESet
SCPI Transition Filters (NTR and PTR registers) none none preset preset
SCPI Enable Registers none none preset preset
SCPI Event Registers none clear clear none
SCPI Error/Event Queue enable none none preset preset
SCPI Error/Event Queue none clear clear none
IEEE488.2 Registers ESE SRE none none clear none
IEEE488.2 Registers SESR STB none clear clear none

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals