EasyManua.ls Logo

Agilent Technologies N1911A

Agilent Technologies N1911A
803 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
17 IEEE 488.2 Command Reference
740 N1911A/1912A P-Series Power Meters Programming Guide
*DDT <arbitrary block program data>|<string program data>
The *DDT (Define Device Trigger) command determines the power meter’s
response to a GET (Group Execute Trigger) message or *TRG common
command. This command effectively turns GET and *TRG into queries, with
the measured power being returned.
Syntax
Parameters
1
The <action> field of the parameter may contain:
FETC?
FETC1?
FETC2? (N1912A only)
*TRG
TRIG1
TRIG2 (N1912A only)
2
The first digit after the # indicates the number of following digits. The following digits indicate
the length of the data.
Type Description Range of Values
arbitrary block program
data
The command which is
executed on a GET or
*TRG.
#nN<action>
1,2
string program data
“<action>”
1
string program data
*DDT
Space
arbitrary block program data
?

Table of Contents

Other manuals for Agilent Technologies N1911A

Related product manuals