OUTput Subsystem 8
N1911A/1912A P-Series Power Meters Programming Guide 301
OUTPut:ROSCillator[:STATe] <boolean>
This command enables or disables the POWER REF output.
Syntax
Example
Reset Condition
On reset, the POWER REF output is disabled.
Query
OUTPut:ROSCillator[:STATe]?
The query command enters a 1 or 0 into the output buffer indicating
whether or not the POWER REF is enabled.
• 1 is returned when the POWER REF output is enabled
• 0 is returned when the POWER REF output is disabled
OUTP:ROSC:STAT 1 This command enables the POWER REF
output.
OUTP :STAT
:ROSC 0|OFF
1|ON