1 Power Meter Remote Operation
62 N1911A/1912A P-Series Power Meters Programming Guide
Tab le 1-7 Bit Definitions - Questionable Status Registers
The condition bits are set and cleared under the following conditions:
Tab le 1-8 Bit change conditions for Questionable Status Register
Bit
Number
Decimal
Weight
Definition
0 to 2 - Not used
38 POWer Summary
4 to 7 - Not used
8 256 CALibration Summary
9 512 Power On Self Test
10 to 14 - Not Used
15 - Not used (always 0)
Bit
Number
Meaning EVENts Causing Bit Changes
3 POWer
Summary
This is a summary bit for the Questionable POWer Register.
SET:
Error –230, “Data corrupt or stale”
Error –231, “Data questionable;Input Overload”
Error –231, “Data questionable;Input Overload ChA
*
Error –231, “Data questionable;Input Overload ChB”
*
Error –231, “Data questionable;PLEASE ZERO”
Error –231, “Data questionable;PLEASE ZERO ChA
*
Error –231, “Data questionable;PLEASE ZERO ChB”
*
Error –231, ”Data questionable;Lower window log error”
*
Error –231, ”Data questionable;Upper window log error”
*
CLEARED: When no errors are detected by the power
meter during a measurement covering the causes given
for it to set.

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals