FORMat Subsystem 6
N1911A/1912A P-Series Power Meters Programming Guide 251
FORMat[:READings]:BORDer <character_data>
This command controls whether the binary data is transferred in normal
or swapped Byte ORDer. It is only used when
FORMat[:READings][:DATA] is set to REAL.
Syntax
Parameters
Example
Reset Condition
On reset, this value is set to NORMal.
Item Description/Default Range of Values
character_data Byte order of binary data transfer:
• NORMal
• SWAPped
NORMal
SWAPped
FORM:BORD SWAP This command sets the byte order to
swapped.
FORM
:READ :BORD
Space
?
character_data