2 MEASurement Commands
156 N1911A/1912A P-Series Power Meters Programming Guide
MEASure[1]|2|3|4[:SCALar][:POWer:AC]:RATio?
[<expected_value>[,<resolution>[,<source list>]]]
This command sets the specified window’s measurement function to ratio
mode with relative mode off, aborts, configures then initiates both
Channel A and B, calculates the ratio measurement result and places the
result on the bus. The result is a ratio based measurement and is
expressed in the units defined by UNIT[1]|2|3|4:POWer:RATio.
Syntax
Parameters
Refer to “Optional Parameters” on page 90 for additional details on the
parameters in this command.
:POW
:AC
?
:SCAL
:RAT
Space
expected_value
DEF
resolution
,
DEF
source list
,
MEAS
1
2
3
4
Item Description/Default Range of Values
expected_value
(for the expected
power level)
The power meter ignores the numeric value
entered in this parameter. Any value entered
is treated like DEF.
sensor dependent
DEF
1
resolution A numeric value for the resolution. If
unspecified the current resolution setting is
used.
1 to 4
2
1.0, 0.1, 0.01, 0.001
DEF
1