10 SENSe Subsystem
436 N1911A/1912A P-Series Power Meters Programming Guide
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:AUTO:REF1|REF2
<numeric_value>
This command is used to set the Reference 1 and 2 of the selected gate
for Auto Gating Marker.
Syntax
Parameters
NOTE
This command is only applicable when N192x or E932x sensor is present and trigger
acquisition mode is selected.
Item Description/Default Range of Values
numeric_value The values of Auto Gating Marker References 1
and 2 for the selected gate.
The combined value of REF1 and REF2 can not
exceed 99.9 %.
0.0 to 99.9
Space
:REF1
?
1
numeric_value
2
4
3
:REF2
:AUTO
SENS
1
SWE
SENS2
: