CALCulate Subsystem 3
N1911A/1912A P-Series Power Meters Programming Guide 167
Query
CALCulate[1]|2|3|4:FEED[1]2?
The query returns the current value of the string.
Query Example
Error Message
If the command is used when no sensor is attached, error –241
“Hardware missing” occurs.
•If <string> contains ON SWEEP[1]|2|3|4 and the feed’s TRIG:SOUR is
not INT or EXT (for single channel power meters) or INT1, INT2 or EXT
(for dual channel power meters), error 221 “Settings conflict” occurs.
If the command changes the measurement mode to PEAK or PTAV when
a sensor other than a P- Series or E9320 power sensor is connected or
a P- Series or E9320 Sensor is connected and set to AVERage mode
rather than NORMal mode, error –221, “Settings Conflict” occurs.
CALC1:FEED2? This command queries the current setting
of the data_handle on FEED2 of the upper
window/upper measurement.