TRIGger Subsystem 14
N1911A/1912A P-Series Power Meters Programming Guide 667
TRIGger[:SEQuence]:HOLDoff <numeric_value>
This command sets the trigger holdoff in seconds.
Syntax
Parameters
Item Description/Default Range of Values
numeric_value The trigger holdoff in seconds.
DEF: the default value is 1 µs
MIN: 1 µs
MAX: 400 ms
Units are resolved to 1 ns.
1 µs to 0.4 seconds
DEF
MIN
MAX
:HOLD
Space
?
numeric_value
DEF
TRIG
:SEQ
MIN
MAX
NOTE
Trigger holdoff is not applicable when the power meter is set to power sweep mode or
frequency sweep mode.