SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 363
[SENSe[1]]|SENSe2:AVERage:COUNt <numeric_value>
This command is used to enter a value for the filter length. If
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO is set to ON then entering a
value for the filter length automatically sets it to OFF. Increasing the value
of filter length increases measurement accuracy but also increases the
time taken to make a power measurement.
Entering a value using this command automatically turns the
[SENSe[1]]|SENSe2:AVERage:STATe command to ON.
NOTE
For most applications, automatic filter length selection
([SENSe[1]]|SENSe2:AVERage:COUNt:AUTO ON) is the best mode of operation.
However, manual filter length selection ([SENSe[1]]|SENSe2:AVERage:COUNt
<numeric_value>) is useful in applications requiring either high resolution or fast
settling times, where signal variations rather than measurement noise need filtering, or
when approximate results are needed quickly.

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals