SERVice Subsystem 15
N1911A/1912A P-Series Power Meters Programming Guide 691
SERVice:BIST:CALibrator <boolean>
This command enables/disables the calibrator self- test during power- up. It
can be used to disable the self- test if it incorrectly indicates a failure. If a
load, for example, a sensor, is connected to the calibrator port this could
cause the self- test to fail. Also, if it fails the self- test, a Pop- up is
displayed for 5 seconds, stating - If Ref Calibrator test fails disconnect any
load attached to it and re- try test.
Syntax
Example
Query
SERVice:BIST:CALibrator?
The query enters a 1 or 0 into the output buffer indicating the status of
the self- test.
• 1 is returned when the self- test is enabled
• 0 is returned when the self- test is disabled
SERV:BIST:CAL OFF This command disables the calibrator
self- test during power- up.
:BIST
SERV 0|OFF
1|ON
Space
?
:CAL
ONCE