14 TRIGger Subsystem
662 N1911A/1912A P-Series Power Meters Programming Guide
TRIGger[1]|2:SOURce
BUS|EXTernal|HOLD|IMMediate|INTernal[[1]|2]
This command configures the trigger system to respond to the specified
source. This command only selects the trigger source. Use the INITiate
command to place the power meter in the wait for trigger state.
Syntax
• This command has been included for compatibility purposes. It has the same purpose as
TRIGger[:SEQuence[1]|2]:SOURce
BUS|EXTernal|HOLD|IMMediate|INTernal[[1]|2] which should be used in
preference.
• If the source is set to INT, INT1 or INT2 when connecting the N1920 or E9320 sensor in
average mode, errors occurs.
TRIG
:SOUR
1
2
BUS
EXT
Space
HOLD
IMM
INT
1
2
?