2 MEASurement Commands
100 N1911A/1912A P-Series Power Meters Programming Guide
CONFigure[1]|2|3|4[:SCALar][:POWer:AC]:RELative
[<expected_value>[,<resolution>[,<source list>]]]
This command sets the measurement function, range and resolution of the
specified window. It sets the measurement function to single channel with
relative mode on. The relative value used is that set by the
CALCulate:RELative:MAGNitude:AUTO command.
Syntax
Parameters
Refer to “Optional Parameters” on page 90 for additional details on the
parameters in this command.
:POW
:AC
:SCAL
:REL
Space
expected_value
DEF
resolution
,
DEF
source list
,
CONF
1
2
3
4
Item Description/Default Range of Values
expected_value
(for the expected
power level)
A numeric value for the expected power
level. The units of measurement are dBm
and W. The default units are defined by
UNIT:POWer.
sensor dependent
DEF
1
resolution A numeric value for the resolution. If
unspecified the current resolution setting
is used.
1 to 4
2
1.0, 0.1, 0.01, 0.001
DEF
1

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals