EasyManuals Logo

Agilent Technologies N1911A Programming Guide

Agilent Technologies N1911A
803 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #475 background imageLoading...
Page #475 background image
SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 447
[SENSe[1]]|SENSe2:TRACe:TIME <numeric_value>
This command sets the duration of the trace for a P- Series sensor and a
E9320 sensors set to NORMal mode. To set the E9320 sensor to NORMal
mode, refer to the command “[SENSe[1]]|SENSe2:DETector:FUNCtion
<character_data>” on page 410.
Syntax
Parameters
Example
Item Description/Default Range of Values
numeric_value The duration of the trace in seconds.
DEF: the default value is 100 µs.
Units are resolved to 1 ns.
20 ns to 1 s
DEF
SENS2:TRAC:TIME 0.5 This command sets the duration of the
trace to 0.5 seconds for Channel B.
SENS
1
SENS2
:
numeric_value
DEF
Space
TRAC
:TIME
?

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals