SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 447
[SENSe[1]]|SENSe2:TRACe:TIME <numeric_value>
This command sets the duration of the trace for a P- Series sensor and a
E9320 sensors set to NORMal mode. To set the E9320 sensor to NORMal
mode, refer to the command “[SENSe[1]]|SENSe2:DETector:FUNCtion
<character_data>” on page 410.
Syntax
Parameters
Example
Item Description/Default Range of Values
numeric_value The duration of the trace in seconds.
DEF: the default value is 100 µs.
Units are resolved to 1 ns.
20 ns to 1 s
DEF
SENS2:TRAC:TIME 0.5 This command sets the duration of the
trace to 0.5 seconds for Channel B.
SENS
1
SENS2
:
numeric_value
DEF
Space
TRAC
:TIME
?