TRIGger Subsystem 14
N1911A/1912A P-Series Power Meters Programming Guide 657
INITiate[:IMMediate]:SEQuence[1]|2
This command sets the power meter in the wait for trigger state. When a
trigger is received, the measurement is taken and the result placed in the
power meter memory. If TRIGger:SOURce is set to IMMediate the
measurement begins as soon as INITiate:IMMediate is executed.
Use FETCh? to transfer a measurement from memory to the output buffer.
Refer to “FETCh[1]|2|3|4 Queries” on page 110 for further information.
Syntax
Example
Error Messages
If the power meter is not in the “idle” state or INITiate:CONTinuous is
ON, error –213, “INIT ignored” occurs.
NOTE
This command performs the same function as INITiate[1]|2:[IMMediate].
INIT:IMM:SEQ1 This command places Channel A in
the wait for trigger state.
INIT
:IMM
:SEQ
1
2