EasyManuals Logo

Agilent Technologies N1911A Programming Guide

Agilent Technologies N1911A
803 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #685 background imageLoading...
Page #685 background image
TRIGger Subsystem 14
N1911A/1912A P-Series Power Meters Programming Guide 657
INITiate[:IMMediate]:SEQuence[1]|2
This command sets the power meter in the wait for trigger state. When a
trigger is received, the measurement is taken and the result placed in the
power meter memory. If TRIGger:SOURce is set to IMMediate the
measurement begins as soon as INITiate:IMMediate is executed.
Use FETCh? to transfer a measurement from memory to the output buffer.
Refer to “FETCh[1]|2|3|4 Queries” on page 110 for further information.
Syntax
Example
Error Messages
If the power meter is not in the “idle” state or INITiate:CONTinuous is
ON, error –213, “INIT ignored” occurs.
NOTE
This command performs the same function as INITiate[1]|2:[IMMediate].
INIT:IMM:SEQ1 This command places Channel A in
the wait for trigger state.
INIT
:IMM
:SEQ
1
2

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals