EasyManua.ls Logo

Agilent Technologies N1911A

Agilent Technologies N1911A
803 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7MEMory Subsystem
276 N1911A/1912A P-Series Power Meters Programming Guide
MEMory:STATe:DEFine <character_data>,<numeric_value>
This command is used to associate a name with a save/recall register
number.
Syntax
Parameters
Example
Query
MEMory:STATe:DEFine? <string>
The query returns the register number for the given register name.
Item Description/Default Range of Values
character_data Details the register name. A maximum of 12
characters can be used.
A to Z (uppercase)
a to z (lowercase)
0-9
_ (underscore)
numeric_value A numeric value (<NRf>) for the register
number.
0 to 9
MEM:STAT:DEF "SETUP1",4 This command names register 4 SETUP1.
MEM
:STAT
numeric_value
:DEF
,
Space
?
Space
character_data
character_data

Table of Contents

Other manuals for Agilent Technologies N1911A

Related product manuals