EasyManua.ls Logo

Agilent Technologies N1911A

Agilent Technologies N1911A
803 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 449
[SENSe[1]]|SENSe2:TRACe:UNIT <character_data>
This command sets the units for the trace for the specified channel.
Syntax
Parameters
Example
Reset Condition
On reset the units are set to dBm.
Item Description/Default Range of Values
character_data DBM: dBm
•W: Watts
DBM
W
SENS2:TRAC:UNIT W This command sets the trace units for
Channel B to Watts.
SENS
:TRAC
Space
character_data
?
1
2
:UNIT

Table of Contents