CALibration Subsystem 4
N1911A/1912A P-Series Power Meters Programming Guide 209
CALibration[1]|2:RCALibration <boolean>
This command enables and disables the zero/cal lockout facility. With the
lockout facility enabled the power meter is stopped from making
measurements until the connected sensor has been zeroed and calibrated.
Syntax
Example
Reset Condition
On reset, the state of the zero/cal lockout is unaffected.
Query
CALibration[1]|2:RCALibration?
The query enters a 1 or 0 into the output buffer indicating whether
zero/cal lockout is enabled or disabled.
• 1 is returned if zero/cal lockout is enabled
• 0 is returned if zero/cal lockout is disabled
CAL1:RCAL 1 This command enables the zero/cal
lockout facility on Channel A.
:RCAL
0|OFF
1|ON
Space
?
CAL
1
2