OUTput Subsystem 8
N1911A/1912A P-Series Power Meters Programming Guide 293
OUTPut:RECorder[1]|2:FEED <data_handle>
This command specifies which measurement is sent to the recorder output
specified by the numeric value following RECorder. RECorder1 applies to
both single and dual channel power meters. RECorder2 applies to dual
channel power meters only.
Syntax
Parameters
Example
Reset Condition
On reset, data_handle is set to its previous value.
Item Description/Default Range of Values
data_handle The CALC block specifying the
measurement to be sent to the recorder
output.
“CALC1” or “CALC”
“CALC2”
“CALC3”
“CALC4”
OUTP:REC2:FEED “CALC1” This command sends the CALC1
measurement to recorder output 2.
OUTP :FEED
:REC
Space
?
1
2
data_handle