EasyManuals Logo

Agilent Technologies N1911A Programming Guide

Agilent Technologies N1911A
803 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #321 background imageLoading...
Page #321 background image
OUTput Subsystem 8
N1911A/1912A P-Series Power Meters Programming Guide 293
OUTPut:RECorder[1]|2:FEED <data_handle>
This command specifies which measurement is sent to the recorder output
specified by the numeric value following RECorder. RECorder1 applies to
both single and dual channel power meters. RECorder2 applies to dual
channel power meters only.
Syntax
Parameters
Example
Reset Condition
On reset, data_handle is set to its previous value.
Item Description/Default Range of Values
data_handle The CALC block specifying the
measurement to be sent to the recorder
output.
“CALC1” or “CALC”
“CALC2”
“CALC3”
“CALC4”
OUTP:REC2:FEED “CALC1” This command sends the CALC1
measurement to recorder output 2.
OUTP :FEED
:REC
Space
?
1
2
data_handle

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals