PSTatistic Subsystem 9
N1911A/1912A P-Series Power Meters Programming Guide 329
PSTatistic:CCDF:REFerence:POWer:PTAVerage?
This command is used to retrieve peak to average data of the saved
reference trace.
Syntax
Example
Error Messages
If command is executed in other window besides CCDF window, error
–221 "Settings conflict: Requires CCDF window" occurs.
If there was no previously saved trace, error –221 "Settings conflict: No
reference trace saved" occurs. Please check the status of saved reference
trace using command PST:CCDF:STOR:REF?.
NOTE
This command is only applicable when P-Series sensors are present and the following
conditions are met:
Free-run acquisition mode is selected
NORMal or DOUBle measurement speed setting is chosen
:POW
PST :CCDF :REF
?
:PTAV
PST:CCDF:REF:POW:PTAV? This command returns the peak to
average power of the saved reference
trace.