Calibration Factor Block Layout A
N1911A/1912A P-Series Power Meters Programming Guide A-765
Tab le A-9 7 Calibration Factor Block Layout: E9300 Series Sensors
These table entries are repeated as shown for each frequency point
Frequency (point ‘N’) 4 - 32 bit fixed 0 to Fhpb*
(2^32)
None Fhbp = Freq Hz per bit
Cal factor (low power)
1
2 - 2.14 0.25 to 3 None Power (in watts) is
divided by this value.
Cal factor (high power)
1
2 - 2.14 0.25 to 3 None Power (in watts) is
divided by this value.
Tab le Siz e :
-
See note
1
The table size is
dependent on the
number of frequency
points.
E4410 Series
Sensors: Calibration
Factor Block Layout
No.
Bytes
Contents Data
Format
Data
Range
Units Notes
E9300 Series
Sensors: Calibration
Factor Block Layout
No.
Bytes
Contents Data
Format
Data
Range
Units Notes
Header:
Number of tables 1 2 None Number of cal factor
tables. Note that the
power levels and
frequency points are the
same for all tables.
Number of frequency
points
2 - 16 bit
integer
None
Bytes per frequency
point
1 - - - None Number of bytes in cal
factor value at each
frequency and power
level.
Frequency LSB weight 2 1000 - Hertz Fhbp (Freq. Hz per bit).
1 KHz per bit for the cal
factor: 1KHz x 2^32 =
4.3E+12 = 4300 GHz
range.

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals