EasyManua.ls Logo

Agilent Technologies N1911A

Agilent Technologies N1911A
803 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 SENSe Subsystem
410 N1911A/1912A P-Series Power Meters Programming Guide
[SENSe[1]]|SENSe2:DETector:FUNCtion <character_data>
This command sets the measurement mode for the E9320 and N1920
sensor
either to average or peak sensor mode.
Syntax
Parameters
1
When measurement mode is set to average:
If TRIG:SOUR is set to INT1, INT2 or EXT, it is set automatically to IMM.
INIT:CONT is set automatically to ON.
SENS:AVER2:STAT is set automatically to OFF.
CALC:FEED is set automatically to “POW:AVG for all CALC blocks using the specified channel in
their CALC:MATH:EXPR.
2
When measurement mode is set to NORMal:
SENS:CORR:DCYC:STAT is set automatically to OFF.
Item Description/Default Range of Values
character_data Defines the measurement mode:
AVERage: sets the E9320 and P-Series
sensor to average only mode.
NORMal: sets the E9320 and P-Series
sensor to normal mode.
AVERage
1
NORMal
2
DET
:FUNC
SENS
1
SENS2
:
Space
?
character_data

Table of Contents

Other manuals for Agilent Technologies N1911A

Related product manuals