10 SENSe Subsystem
410 N1911A/1912A P-Series Power Meters Programming Guide
[SENSe[1]]|SENSe2:DETector:FUNCtion <character_data>
This command sets the measurement mode for the E9320 and N1920
sensor
either to average or peak sensor mode.
Syntax
Parameters
1
When measurement mode is set to average:
If TRIG:SOUR is set to INT1, INT2 or EXT, it is set automatically to IMM.
INIT:CONT is set automatically to ON.
SENS:AVER2:STAT is set automatically to OFF.
CALC:FEED is set automatically to “POW:AVG for all CALC blocks using the specified channel in
their CALC:MATH:EXPR.
2
When measurement mode is set to NORMal:
SENS:CORR:DCYC:STAT is set automatically to OFF.
Item Description/Default Range of Values
character_data Defines the measurement mode:
AVERage: sets the E9320 and P-Series
sensor to average only mode.
NORMal: sets the E9320 and P-Series
sensor to normal mode.
AVERage
1
NORMal
2
DET
:FUNC
SENS
1
SENS2
:
Space
?
character_data

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals