SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 411
Example
Reset Condition
On reset, the mode is set to NORMal.
Query
[SENSe[1]]|SENSe2:DETector:FUNCtion?
The query returns the current sensor mode setting.
Query Example
Error Messages
If the command is used when a non E9320 or N1920 sensor is
connected, the error –241, “Hardware missing” occurs.
SENS1:DET:FUNC NORM This command sets the sensor to peak
mode for Channel A.
SENS:DET:FUNC? This command queries the current sensor
mode setting for Channel A.