SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 445
[SENSe[1]]|SENSe2:TRACe:OFFSet:TIME <numeric_value>
This command sets the delay between the delayed trigger point and the
start of the trace for P- Series or E9320 sensors are set to NORMal mode.
To set an E9320 sensor to NORMal mode, refer to the command
“[SENSe[1]]|SENSe2:DETector:FUNCtion <character_data>” on page 410.
Syntax
Parameters
Example
Item Description/Default Range of Values
numeric_value The length of the delay in seconds.
DEF: the default value is 0 seconds
Units are resolved to 1 ns.
–1 to 1 second
DEF
SENS:TRAC:OFFS:TIME TIME
0.05
This command sets the delay to 0.05
seconds.
SENS
1
:TIME
SENS2
:
numeric_value
DEF
Space
TRAC
?
:OFFS