2 MEASurement Commands
98 N1911A/1912A P-Series Power Meters Programming Guide
CONFigure[1]|2|3|4[:SCALar][:POWer:AC]
[<expected_value>[,<resolution>[,<source list>]]]
This command is used on the specified window/measurement to set:
• The expected power level of the measurement.
• The resolution of the window/measurement.
• The channel on which the measurement will be made.
Syntax
Parameters
Refer to “Optional Parameters” on page 90 for additional details on the
parameters in this command.
Space
expected_value
DEF
resolution
,
:POW
:AC
:SCAL
DEF
source list
,
CONF
1
2
3
4
Item Description/Default Range of Values
expected_value A numeric value for the expected power
level. The units of measurement are dBm
and W. The default units are defined by
UNIT:POWer.
Sensor dependent.
DEF
1
resolution A numeric value for the resolution. If
unspecified the current resolution setting is
used.
1 to 4
2
1.0, 0.1, 0.01, 0.001
DEF
1