EasyManua.ls Logo

Agilent Technologies N1911A

Agilent Technologies N1911A
803 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2 MEASurement Commands
98 N1911A/1912A P-Series Power Meters Programming Guide
CONFigure[1]|2|3|4[:SCALar][:POWer:AC]
[<expected_value>[,<resolution>[,<source list>]]]
This command is used on the specified window/measurement to set:
The expected power level of the measurement.
The resolution of the window/measurement.
The channel on which the measurement will be made.
Syntax
Parameters
Refer to “Optional Parameters” on page 90 for additional details on the
parameters in this command.
Space
expected_value
DEF
resolution
,
:POW
:AC
:SCAL
DEF
source list
,
CONF
1
2
3
4
Item Description/Default Range of Values
expected_value A numeric value for the expected power
level. The units of measurement are dBm
and W. The default units are defined by
UNIT:POWer.
Sensor dependent.
DEF
1
resolution A numeric value for the resolution. If
unspecified the current resolution setting is
used.
1 to 4
2
1.0, 0.1, 0.01, 0.001
DEF
1

Table of Contents

Other manuals for Agilent Technologies N1911A

Related product manuals