UNIT Subsystem 16
N1911A/1912A P-Series Power Meters Programming Guide 733
UNIT[1]|2|3|4:POWer <amplitude_unit>
This command sets the power measurement units for a specified
window/measurement. The power suffix set by UNIT:POWer is used for any
command which accepts a numeric value in more than one unit
For the N1911A:
• UNIT1:POWer sets the power measurement units for the upper
window/upper measurement.
• UNIT2:POWer sets the power measurement units for the lower
window/upper measurement.
• UNIT3:POWer sets the power measurement units for the upper
window/lower measurement.
• UNIT4:POWer sets the power measurement units for the lower
window/lower measurement.
For ratio and relative power measurements:
•If UNIT:POWer is W, the measurement units are percentage.
•If UNIT:POWer is DBM, the measurement units are dB relative.
Syntax
amplitude_unit
:POW
Space
?
UNIT
1
2
3
4