2 MEASurement Commands
148 N1911A/1912A P-Series Power Meters Programming Guide
MEASure[1]|2|3|4[:SCALar][:POWer:AC]?
[<expected_value>[,<resolution>[,<source list>]]]
This command sets the specified window’s measurement function to single
channel with relative mode off, aborts, configures the window then
initiates Channel A or B, calculates the measurement result and places the
result on the bus.
Syntax
Parameters
Refer to “Optional Parameters” on page 90 for additional details on the
parameters in this command.
:POW
:AC
:SCAL
?
Space
expected_value
DEF
resolution
,
DEF
source list
,
MEAS
1
2
3
4
Item Description/Default Range of Values
expected_value
(for the expected
power level)
A numeric value for the expected power level.
The units of measurement are dBm and W.
The default units are defined by
UNIT:POWer.
sensor dependent
DEF
1
resolution A numeric value for the resolution. If
unspecified the current resolution setting is
used.
1 to 4
2
1.0, 0.1, 0.01, 0.001
DEF
1