15 SERVice Subsystem
696 N1911A/1912A P-Series Power Meters Programming Guide
SERVice:BIST:PEAK[1]|2:LINearity <numeric_value>
This command initiates the PEAK linearity test.
Syntax
Parameters
Example
Item Description/Default Range of Values
Numeric_value Define the number of samples taken for
results, default:0
0 to 8000
SERV:BIST:PEAK:LIN 8000 This command sets the number of samples
of the PEAK linearity test to be 8000.
:BIST
1
:PEAK
2
SERV
:LIN
Space
?
numeric_value
DEF

Table of Contents