TRIGger Subsystem 14
N1911A/1912A P-Series Power Meters Programming Guide 671
TRIGger[:SEQuence]:LEVel <numeric_value>
This command sets the power level at which a trigger event is recognized.
Syntax
Parameters
1
If a channel offset has been previously set, a higher numeric value is permitted. See“Setting
Offsets” on page 38 for more information.
Item Description/Default
Range of Values
1
numeric_value The power level at which a trigger event is
recognized.
DEF: the default value is 0 dBm
Units are resolved to 0.1 dBm.
–40 to 20 dBm
DEF
:LEV
Space
numeric_value
?
DEF
TRIG
:SEQ