SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 397
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3 Commands
These commands control the pulse power measurement feature of the
power meter.
The following commands are detailed in this section:
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3[:INPut]
[:MAGNitude] <numeric_value>
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3:STATe <boolean>
You can use either DCYCLe or GAIN3 in these commands, both do the same. Using
GAIN3 complies with the SCPI standard whereas DCYCle does not, but may make your
program more understandable.