EasyManuals Logo

Agilent Technologies N1911A Programming Guide

Agilent Technologies N1911A
803 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #424 background imageLoading...
Page #424 background image
10 SENSe Subsystem
396 N1911A/1912A P-Series Power Meters Programming Guide
Query
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe?
The query returns a 1 or 0 into the output buffer indicating whether a
table is enabled or disabled.
1 is returned when the table is enabled
0 is returned when the table is disabled
Query Example
Error Messages
If you attempt to set this command to ON and no table has been
selected using
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:[SELect] then error
–221, “Settings conflict” occurs and
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe remains OFF.
If the CSET1 command is used when a P- Series, N8480 Series
(excluding Option CFT) or an E-Series power sensor is connected, the
error –241 “Hardware missing” occurs.
SENS2:CORR:CSET1:STAT? This command queries whether there is
currently an active sensor calibration
table for Channel B.

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals