EasyManuals Logo

Agilent Technologies N1911A Programming Guide

Agilent Technologies N1911A
803 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #138 background imageLoading...
Page #138 background image
2 MEASurement Commands
110 N1911A/1912A P-Series Power Meters Programming Guide
FETCh[1]|2|3|4 Queries
The FETCh? queries set the specified window’s measurement function. This
can be set to either single channel, difference or ratio measurements, with
relative mode either off or on. They then recalculate the measurement and
place the result on the bus. The format of the result is set by
FORM[:READ][:DATA]. Refer to Chapter 6, “FORMat Subsystem,” on page
249 for further information.
The query returns a measurement result when it is valid. The
measurement result is invalid under the following conditions:
•When *RST is executed.
Whenever a measurement is initiated.
•When any SENSe parameter, such as frequency, is changed.
If data is invalid, the FETCh? query is not completed until all data
becomes valid. The exceptions to this are, if the power meter is in the idle
state and the data is invalid, or the power meter has been reconfigured as
defined above and no new measurement has been initiated. In such cases,
the FETCh? routine generates the error –230, “Data corrupt or stale” and
no result is returned. A common cause for this error is receiving a FETCh?
after a *RST. If the expected value and resolution parameters are not the
same as those that were used to collect the data, error –221, “Settings
conflict” occurs.
NOTE
When TRIG:SOUR is INT1, INT2 or EXT and a new acquisition has been initiated (using the
INIT command for example), FETCH? waits until the trigger takes place before executing. If triger
conditions are not satisfied - when the trigger level differs greatly from the signal level for example -
this can give the impression that the power meter has hung.
To unlock the power meter and adjust trigger settings, an SDC (Selected Device Clear) GPIB
Command must be performed. This is equivalent to “EXECUTE CLEAR” in Agilent VEE.

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals