OUTput Subsystem 8
N1911A/1912A P-Series Power Meters Programming Guide 299
OUTPut:RECorder[1]|2:STATe <boolean>
This command enables or disables the specified recorder output.
Syntax
Example
Reset Condition
On reset, the recorder output is OFF.
Query
OUTPut:RECorder[1]|2:STATe?
The query command enters a 1 or 0 into the output buffer indicating
whether or not the specified recorder is switched on.
• 1 is returned when the recorder output is switched ON
• 0 is returned when the recorder output is switched OFF
OUTP:REC1:STAT 1 This command enables the specified
recorder output.
OUTP
:STAT
:REC
0|OFF
1|ON