TRIGger Subsystem 14
N1911A/1912A P-Series Power Meters Programming Guide 659
TRIGger[1]|2:DELay:AUTO <boolean>
This command is used to determine whether or not there is a
settling- time delay before a measurement is made.
When this command is set to:
• ON, the power meter inserts a settling- time delay before taking the
requested measurement. This settling time allows the internal digital
filter to be updated with new values to produce valid, accurate
measurement results. The trigger with delay command allows settling
time for the internal amplifiers and filters. It does not allow time for
power sensor delay.
In cases of large power changes, the delay may not be sufficient for
complete settling. Accurate readings can be assured by taking two
successive measurements for comparison.
• OFF, the power meter makes the measurement immediately a trigger is
received.
TRIGger[1]|2:DELay:AUTO is ignored if TRIGger[1]|2[:IMMediate] is
set to ON.
Syntax
TRIG
:AUTO
:DEL
0|OFF
1|ON
Space
?
1
2
NOTE
Trigger delay is not applicable when the power meter is set to power sweep mode or
frequency sweep mode.

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals